This error occurs when the "Definition License" is not registered on the dedicated SD card.
| Cause | Detail |
| License not installed | The necessary license files (*.LCT or *.ALC) are not installed on the dedicated SD card. |
| The definition and license do not match. | The definition file (*.CM file) you are trying to use does not match the license currently registered on the SD card. |
| Copying and using the YIM folder | If you copy and use a YIM folder (containing the loaded definition files) created on another SD card, the necessary additional licenses will not be registered on the destination SD card. |
<countermeasure>
- To view the license registered on the dedicated SD card, click [Read License] in the "License Information" tab of the SWX600.
- Please register your product (definition license) on the support webpage and obtain the license file (.LCT, etc.).
- Click "Add License" from the "License Information" tab on the SWX600 and add the acquired license file to the dedicated SD card.
*If you encounter an error (1017: PROBE LOGIC NO LICENCE) related to the probe logic function license (probe logic license), you need to add the license to the "NETIMPRESS main unit" rather than the SD card, but the license registration method is the same.
* For NETIMPRESS air, you can perform "Read License" or "Add License" from the "air Information" tab in NETIMPRESS air Connect (AZ990).