1. Meaning of the Error
This error occurs when there is a communication or control failure between the
programmer main unit (AFX100 / AFX210) and the CAN probe hardware (PHX410).
This error may also be detected when there is no response from the target MCU.
This error occurs when there is a communication or control failure between the
programmer main unit (AFX100 / AFX210) and the CAN probe hardware (PHX410).
This error may also be detected when there is no response from the target MCU.
2. Causes and Countermeasures
■ Cause 1: Connection failure or disconnection of the cable between the
programmer main unit and the probe hardware
-
Countermeasure:
-
Disconnect the M12 cable between the programmer and the probe, and
reconnect it securely. - If you have a spare M12 cable, replace the cable to see if the symptom improves.
-
Note: Bending the M12 cable sharply may cause the internal wiring to break.
Please be careful when routing the cable.
-
Disconnect the M12 cable between the programmer and the probe, and
■ Cause 2: Hardware failure
-
Countermeasure:
-
A physical failure of the probe hardware (PHX410) or the programmer main unit
(AFX100 / AFX210) is suspected. -
If you have spare equipment, individually replace the probe hardware and the
programmer main unit to identify the location of the failure (isolate the cause).
-
A physical failure of the probe hardware (PHX410) or the programmer main unit
■ Cause 3: Connection failure with the target
-
Countermeasure:
-
Recheck the wiring on the jig side (for poor contact or disconnection) and the
connection status to the target board.
-
Recheck the wiring on the jig side (for poor contact or disconnection) and the
■ Cause 4: Communication error due to the effect of common mode noise via the
AC line
AC line
-
Countermeasure:
-
Attach a ground wire to the ground terminal of the PHX410 and ensure proper
grounding. -
Insert a noise filter, such as a noise-cut transformer, into the AC line as a
countermeasure.
-
Attach a ground wire to the ground terminal of the PHX410 and ensure proper
■ Cause 5: Programmer setting or connection issue
-
Countermeasure:
-
Check if the "Probe Select (Probe1 / Probe2)" setting in the software correctly
matches the channel actually connected. -
Make sure that a serial communication probe (PHX400) is not mistakenly
connected to the selected channel. Please connect the CAN communication
probe (PHX410) correctly.
-
Check if the "Probe Select (Probe1 / Probe2)" setting in the software correctly