[Supported model]
adviceLUNA, adviceLUNA2
[Scope of application]
ARM
[Error message]
ICE Error No.90D: Time-out error waiting for break.
[Workaround]
- Check if an MPU-specific file (xxx.mpv) is matching with the MPU of a user system.
- Check nSRST pin and nTRST pin of JTAG are connected in accordance with the description of ”Notes and points for connecting with a user system” in the adviceLUNAII User’s Manual.
- When using an MPU of software macro core, choose [MPU] - [MPU-Specific Settings] from the menu bar of the microVIEW-PLUS.
And check on Adaptive Clock on the [User System] tab of the [MPU-Specific Settings] dialog box.
- MPU may be runaway.
Check the program executed from reset vector is correct.
In the case there is no program, turn the user system power on or press the user system reset button.
- MPU may sleep.
MPU does not wake up even there is a request to wake up by ICE.
Check that there is no forced off of MPU power or forced stop of MPU clock by user system.
[Cause]
This error is output when there is no response for a certain time when a break is invoked to a user system by the ICE.
[Caution]
In the case this error occurred, re-initialization of ICE and user system may require.
For recovery steps, see Chapter 25.1 ”Re-initialization of User System” (adviceLUNA).
For recovery steps, see Chapter 26.1 ”Re-initialization of User System” (adviceLUNA2).