[Supported model]
adviceLUNA, adviceLUNA2
[Scope of application]
microVIEW-PLUS
[Error message]
Error No.911: Invalid Group Name
[Workaround]
Correct the description of the register group name while referring to Section 10.8 “Description of the User Register File”.
[Cause]
This error message appears if no“]”is found, the length of the character string is 32 bytes or more or the register group name includes“,”,“[”“]” (excluding the first and last “[”,“]”) when reading the user defined register file (∗.uer).